0086-550-7096433
    Disclaimer|Sitemap| Contact
1.jpg
  • 1.jpg

Monocrystalline silicon pressure transmitter

Monocrystal silicon pressure transmitter has 0.075% ultra-high measurement accuracy, 100:1 wide rang…
Description

Monocrystal silicon pressure transmitter has 0.075% ultra-high measurement accuracy, 100:1 wide range ratio, high stability, fast response speed, high reliability, with HART communication function convenient manual debugging, with intrinsically safe, flameproof type, optional a variety of isolation diaphragm structure, a variety of process connection standards, Suitable for pressure measurement under micropressure, absolute pressure, negative pressure, high temperature and high pressure, low temperature and corrosion conditions.


Working principle of monocrystalline silicon pressure transmitter


Using the piezoresistive effect of monocrystalline silicon. The monocrystalline silicon wafer is used as the elastic element, and a set of equivalent resistance is diffused in the specific direction of the monocrystalline silicon wafer by the process of integrated circuit on the monocrystalline silicon membrane, and the resistance is connected into a bridge, and the monocrystalline silicon wafer is placed in the sensor cavity. When the pressure changes, the monocrystalline silicon produces strain, which causes the strain resistance directly diffused on it to produce changes proportional to the measured pressure, and then the corresponding voltage output signal is obtained by the bridge circuit.


Two, monocrystalline silicon pressure transmitter technical parameters


Measuring range

Relative pressure: 0 ~ 400Pa ~ 35MPa


J.D pressure: 0 ~ 1.75kpa ~ 35MPa

Signal output

4-20MA +HART Communication protocol (two-wire system)


Accuracy of measurement

General error: 0.1% (including linearity, repeatability and hysteresis), 0.075% for special requirements


Long-term stability

Better than 0.15% F.S. per year

Five years better than 0.2%F.S

Temperature range


Allowable medium temperature :(-40 ~ 120) ℃


Allowable ambient temperature :(-40 ~ 85) ℃


Permissible storage temperature :(-40 ~ 85) ℃


The temperature influence

Relative to 25℃ : ≤±0.4% standard range


Damping time

0~40S continuous adjustable


Vibration resistance: 2g 1000HZ ±0.05%g


Impact resistance: 50G 11MESC


Electromagnetic compatibility: comply with IEC801 standard


Sensor allows overvoltage: standard range 3 times pressure


Zero z.uni small alarm current range: 3.8mA


Full degree Z.ui large alarm current range: 21mA


Start time: 3S without preheating


Part of the material in contact with the medium


Process connector: stainless steel 316, Hastelloy C


Diaphragm of sensor: Stainless steel 316, Hastelloy C


Part of the material that is not in contact with the medium


Case: die-cast aluminum with epoxy coating


Sensor filling fluid: silicone oil anaerobic oil


Field display: LCD 3 1/2 precision digital display LCD analog strip display


Protection level :IP66


Weight: about 1Kg


Ex D ⅱ C T5, intrinsically safe Ex IA ⅱ C T5


Allowed ambient humidity :100%RH


Browse:
Return To The Top